Search

Neck/Cervical

Abdomen

MR Stroke Protocol (Version 8.1)

This MR stroke protocol allows identification of the stroke patients' histological and anatomical characteristics including vessels from 6 imaging contrasts in almost 6 minutes.

Axial DWI with Exsper

Image
SequenceSEEPI2D
Parallel AcquisitionExsper 2x1
Scan time0:38
Resolution1.6x1.6 mm
FOV23 cm
Matrix size144x144
Slice thickness5 mm
Slice gap1 mm
Number of slices25
TR4130 ms
TE77 ms
NAQ1
b-value1000

Axial FLAIR with Exsper

Image
SequenceFSE2D
Parallel AcquisitionExsper 2x1
Scan time1:40
Resolution1.0x1.0 mm
FOV23 cm
Matrix size224x224
Slice thickness5 mm
Slice gap1 mm
Number of slices25
TR10000 ms
TE104 ms
NAQ1

Axial TOF MRA with SPEEDER, Fast 3D mode 50%

Image
SequenceMRA3D
Parallel AcquisitionSPEEDER 2x1
Scan time1:15
Resolution1.3x0.8 mm
FOV20 cm
Matrix size160x256
Slice thickness1.3 mm
Number of slabs3
Number of slices116
TR14 ms
TE6.5 ms
NAQ1

Axial T2*w with Exsper

Image
SequenceFE2D
Parallel AcquisitionExsper 2x1
Scan time0:53
Resolution1.0x1.0 mm
FOV23 cm
Matrix size240x240
Slice thickness5 mm
Slice gap1 mm
Number of slices25
TR380 ms
TE9 ms
NAQ1

Axial T2w with Exsper

Image
SequenceFSE2D
Parallel AcquisitionExsper 2x1
Scan time0:40
Resolution0.9x0.9 mm
FOV23 cm
Matrix size256x256
Slice thickness5 mm
Slice gap1 mm
Number of slices25
TR4400 ms
TE90 ms
NAQ1

Axial T1w with Exsper

Image
SequenceSE2D
Parallel AcquisitionExsper 2x1
Scan time0:56
Resolution1.0x1.0 mm
FOV23 cm
Matrix size224x224
Slice thickness5 mm
Slice gap1 mm
Number of slices25
TR428 ms
TE8 ms
NAQ1

1.5T System

Vantage Elan / Active Edition (V8.1)

Image

Coil

Octave SPEEDER Head 1.5T

Image

Clinical solution

Advanced intelligent Clear-IQ Engine (AiCE)

AiCE is the world’s first Deep Learning Reconstruction technology for MRI. AiCE intelligently removes noise from images which can improve SNR and enable increased resolution across all applications.

 

Fast 3D mode (Fast 3D TOF: Vantage Elan Product Page)

You can now apply FAST 3D to TOF contrast MRA procedures with the ability to reduce scan times by up to 50%*, while maintaining excellent image consistency.

Neck/Cervical

Brachial Plexus Protocol (Version 8.1)

This Protocol allows to identify the travel of the brachial Plexus to evaluate the level of the damage for the patient with neurological disorders.

SequenceFASE3D
Parallel AcquisitionSPEEDER 2x1
Scan time5:51
Resolution1.3x1.3 mm
FOV43x34 cm
Matrix size336x256
Slice thickness2 mm
Number of slices50
TR3250
TE369
NAQ2

1.5T System

Vantage Elan / Active Edition (V8.1)

Image

Coil

Octave SPEEDER Head 1.5T

Image

Octave SPEEDER Spine 1.5T

Image

Atlas SPEEDER Body 1.5T

Image

Clinical solution

Advanced intelligent Clear-IQ Engine (AiCE)

AiCE is the world’s first Deep Learning Reconstruction technology for MRI. AiCE intelligently removes noise from images which can improve SNR and enable increased resolution across all applications.

Abdomen

MRCP Protocol (Version 8.1)

This Protocol allows to visualize pancreatic and hepatobiliary systems to identify the lesions non-invasively.

SequenceFASE2D
Parallel AcquisitionExsper 2x1
Scan time0:03
Resolution1.5x1.1 mm
FOV38x28 cm
Matrix size256x256
Slice thickness60 mm
Number of slices1
TR3000
TE1000
NAQ1
SequenceFASE3D
Parallel AcquisitionSPEEDER 2x1
Scan time0:20
Resolution1.0x1.0 mm
FOV32x25.6 cm
Matrix size320x256
Slice thickness3.5 mm
Number of slices20
TR2850
TE800
NAQ1

1.5T System

Vantage Elan / Active Edition (V8.1)

Image

Coil

Atlas SPEEDER Body 1.5T

Image

Octave SPEEDER Spine 1.5T

Image

Clinical solution

Advanced intelligent Clear-IQ Engine (AiCE)

AiCE is the world’s first Deep Learning Reconstruction technology for MRI. AiCE intelligently removes noise from images which can improve SNR and enable increased resolution across all applications.

Exsper (Vantage Elan Product Page)

Exsper is a unique parallel imaging technique which provides accelerated exam. Exsper suppresses unfolding errors even in the small FOV imaging.